AOC-20s vial returning error [Error Code:6019]
Cause¶
The following causes are possible.
- Gripper could not grip the vial when returning it and dropped the vial at the handover position.
Solution¶
Check and perform the following.
-
Redo the teaching using the vial to be used for actual analysis.
-
In case of AOC-30i
AOC-30i_AOC-20sU_SamplerTeaching -
In case of AOC-20i
AOC-20i_AOC-20s_SamplerTeaching
-
-
Perform a transport test (in case of AOC-30i).
AOC-30i_AOC-20sU_TransportTest