GC-2060

GC-2060 Maintenance Menu

Click on the area subject to maintenance.

Other Maintenance Menu

 

Check that the column oven temperature is 40 °C or lower.
Check that the temperatures of the injection port and detector are 50 °C or lower.
Performing the work at a high temperature may cause burns. In addition, parts may be damaged if the work is performed at a high temperature.

Stop the gas supply by following the procedure described in the Instruction Manual.
If the work is performed while gas is flowing, high-pressure gas may be discharged, which can cause an accident or damage the instrument.

When replacing parts, use the items that are described in the Instruction Manual and Maintenance Help.
If you use a part that is not listed, that part may be damaged and may not operate properly.
It may also cause injury or malfunction.