MMI Replacement of O-ring and Liner

The Direct liner is also replaced using the same procedure.

Preparation

First, set the instrument to a state in which it can be serviced (Maintenance Mode).

Required devices

No.
Parts name
Category
Required number
Note
①
Tweezers
1
②
Liner
1
Various optional parts are available.
③
O-ring for MMI
1
Optional product with high heat resistance is available.
④
Removal Tool
1
Optional part
⑤
Spacer for MMI
1
Required when using liquid nitrogen to cool MMI and setting it to -50°C or below.

Procedures

Step 1

GC-2060

Open the INJ cover (★1).

Step 2

Turn the liner fixing nut (★2) counterclockwise by hand to remove it.

Step 3

Lift the injection port (★3) slowly.

Note

To prevent damage to the liner, lift the injection port straight up slowly with your left hand.

Step 4

Use tweezers (①) to pull out the old liner (②).

Note

Be careful not to damage the O-ring seating surface of the sample injection port.

Hint

Using the removal tool (④) makes it easy to pull out.

Step 5

Attach the O-ring (③) to the new liner (②).

Hint
  • Temporarily fix the O-ring about 4 mm from the top of the liner, set it into the sample injection port, and press it down with tweezers until it reaches the bottom.

  • The standard accessory liner comes pre-packed with silica wool at the specified position. When using a liner without pre-packed silica wool, pack silica wool according to Guide for Packing Position of Silica Wool .

Step 6

Install the liner (②) into the injection inlet.

Press it down with tweezers until it reaches the bottom.

Step 7

Install the liner fixing nut (★2) so that it meshes securely with the groove of the injection port, and turn it clockwise to fix it.

Hint

Fix the protrusion of the liner fixing nut (★4) into the hole of the MMI cover (★5) before turning the nut to make tightening easier.

Note

Do not overtighten the nut as this may cause the O-ring to be pushed inward, making it impossible to remove the liner.

Note

When setting the MMI cooling to below -50°C using liquid nitrogen, use the spacer (⑤) and tighten the nut further.

Step 8

Close the INJ cover.

Step 9

Exit maintenance mode using the following procedure according to the relevant instrument, and reset the consumable usage counter.