Replacement of Septum and Liner (OCI-NX)

Preparation

First, set the instrument to a state in which it can be serviced (Maintenance Mode).

Required devices

No.
Parts name
Category
Required number
Note
①
Tweezers
1
②
OCI Liner (for OCI-2030 NX/OCI-NX U)
1
Be sure to use the dedicated insert for OCI-2030 NX/OCI-NX U.
③
Flat-Head Screwdriver
1
④
Septum
1
A septum with high heat resistance (HT type) is recommended for OCI.

Procedures

Step 1

Open the INJ cover (★1).

Step 2

Remove the septum nut ASSY (★2) by turning it counterclockwise, and remove the HT septum (④).

Step 3

Using flat-head screwdriver (③), turn the OCI-2030 NX / OCI-NX U liner (②) counterclockwise, and remove it with tweezers (①).

Step 4

Using tweezers (①), insert a new OCI-2030 NX / OCI-NX U liner (②), and turn it with flat-head screwdriver (③) clockwise to fix it.

Note

Stop turning the flat-head screwdriver when the liner reaches the bottom. Over-tightening may cause seizure.

Step 5

Attach the septum (④), fully tighten the septum nut ASSY (★2) clockwise, and return it one half-turn.

Step 6

Close the INJ cover.

Step 7

Follow one of the procedures below to exit maintenance mode and reset the consumable counter.