Installation of Capillary Column(Graphite Vespel Ferrule)

Required devices

No.
Parts name
Category
Required number
Note
①
Spanner (1/4inch×5/16inch)
1
②
Capillary Cutter (Plate Type)
1
Option
③
NUT, SSNE16/012
1
For Graphite Vespel ferrule
④
Graphite Vespel Ferrule
1
⑤
Column Installation Jig (for INJ)
1
⑥
Septum
1
For positioning. Used septum is acceptable.

Procedures

Hint

When performing Direct analysis, refer to Column Installation for Direct Analysis

Step1

Stop the instrument. Then, on the instrument monitor or workstation, check that the temperature of the column oven is lower than 40 °C, check that the temperatures of the injection port and detector are 50 °C or lower, and check that the gas supply is stopped and there is no residual pressure.

Step2

Draw the column through the septum (⑥)

Hint

The septum is used for positioning the ferrule. Used septum is acceptable.
The septum will be removed in a later step. Use a cutter to cut open the septum.

Step3

Cut the end of column with capillary cutter (②) etc.

Hint

After drawing through the septum, cut the end of the column to remove remaining septum.

Step4

Draw the column through the nut (③) and graphite Vespel ferrule (④).

Note

The tapered surface of the graphite Vespel ferrule (④) should be on the nut side.

Step5

Pass the column through the column installation jig (for INJ) (⑤).

Hint

・Adjust the column so that the end of the column is aligned with the end of the jig.
・The length of the inserted column for SPL/MMI/SPI should be "40 mm".

Step6

Tighten the nut (③) and the column installation jig (for INJ) (⑤) with a spanner(①) while fixing the septum.

Note

If the position of the septum is not fixed, the length of the inserted column will vary.

Step7

Remove the column installation jig (for INJ) (⑤).

Step8

Pull the column oven door latch toward you to open the column oven door.

Step9

Hang the column on the column hanger.

Hint

The column hanger is self-standing. Adjusting the number of coils, etc., can be done outside the oven.

Step 10

Install the column to the injection port.

Step11

Tighten the nut (③) by hand, and tighten the nut with the spanner for column nut, 5/16 x 1/4 (①) one turn.
Remove the positioning septum.

Note

Be careful so that the graphite Vespel ferrule and nut do not become detached from the column.
Be careful not to move the positioning septum

Hint

Tighten the nut with the positioning septum lightly pressed against the nut.

Step12

Close the column oven door.

Step13

Start the GC.

Note

If the graphite Vespel ferrule is new, increase the temperature of the column oven to the temperature used in the analysis, cool the column oven, and additionally tighten the nut on the injector side.
If the ferrule is new, leakage may occur at the ferrule when increasing or decreasing the column oven temperature.