Replacing the Syringe (When using the black plunger holder 2)

Plunger holder 2 (black) is an improved part that makes it easier to replace the syringe.
Please use it according to the following procedure.

Video (Removal of AOC Syringe)

Video (Installation of the AOC Syringe)

Required devices

No.
Parts name
Category
Required number
Note
①
Syringe for AOC
1
②
Barrel Holder (5 pcs)
1
③
AOC Plunger Holder 2, Black, 5 pieces
1

Procedures

Step 1

Open the front door.

Step 2

For AOC-20i, press the [STOP] key.

The hold of the plunger motor of the AOC-20i will be released.
For AOC-30i, the hold of plunger motor is automatically released when the door is opened.

Step 3

Unlock the syringe lock, and while holding the knob, remove the syringe.

Step 4

Check for clogging of the syringe and freedom of movement of the plunger.

Clean or replace it as needed.
See Maintenance of Syringe .

Step 5

Attach the barrel holder to the new syringe.

Step 6

The syringe is set in the syringe drive unit by fitting the syringe plunger into the plunger holder.

Step 7

Fix the syringe with the syringe lock.

Step 8

Make sure the plunger tip is at the bottom. If there is some distance between the syringe plunger and the bottom of the syringe barrel, adjust the zero point using the following procedure.

Hint
The end of the plunger is located at Zero point. The end of the plunger is located above Zero point.

Step 9

Loosen the knurled screw (★1) that fixes the syringe plunger in place.

Step 10

Press the [Reset] key.

Move the plunger driving unit to the home position.

Step 11

Press down on the plunger holder lightly, ensure that the end of the plunger is located at Zero point, and tighten plunger fixing knurled screws to fix it in place.

Step 12

Make sure the plunger tip is at the bottom.

Hint
The end of the plunger is located at Zero point. The end of the plunger is located above Zero point.

Step 13

Close the front door of the AOC.

Step 14

Reset the consumable usage counter.

Hint

See the procedures below for details on how to reset the consumable usage count.