Not execute this operation [Error Code:9061]

Cause

The following items may be the cause. Check the following items.

Solution

Operations such as Ignition/illumination were executed while the GC stopped.

The following operation was executed while the GC stopped.

Press [Start GC] on [GC Start Sequence] screen before executing the listed operations.

Operations such as standard diagnosis were executed while the GC started.

The following operation was executed while the GC started.

Press [Stop GC] on [GC Stop Sequence] screen before executing the listed operations.

Time program setting was changed while time program was running.

Change the setting after time program is completed.

Pre-Run program setting was changed while pre-run program was running.

Change the setting after pre-run program is completed.

AOC control was attempted when the instrument and AOC were not connected.

Check the following items.

Calibration was performed in a situation where calibration was not possible.

Check the following items before calibration.