Not execute this operation [Error Code:9061]
Cause¶
The following items may be the cause. Check the following items.
- Operations such as Ignition/illumination were executed while the GC stopped.
- Operations such as standard diagnosis were executed while the GC started.
- Time program setting was changed while time program was running.
- Pre-Run program setting was changed while pre-run program was running.
- AOC control was attempted when the instrument and AOC were not connected.
- Calibration was performed in a situation where calibration was not possible.
Solution¶
Operations such as Ignition/illumination were executed while the GC stopped.¶
The following operation was executed while the GC stopped.
- Ignition/Illumination
- FTD Feedback
- Start analysis
- Batch
- Clean Up
- PREP RUN
- Inj Maintenance
Press [Start GC] on [GC Start Sequence] screen before executing the listed operations.
Operations such as standard diagnosis were executed while the GC started.¶
The following operation was executed while the GC started.
- Standard Diagnosis
- Carrier gas leak check
- Offset calibration of flow controller
Press [Stop GC] on [GC Stop Sequence] screen before executing the listed operations.
Time program setting was changed while time program was running.¶
Change the setting after time program is completed.
Pre-Run program setting was changed while pre-run program was running.¶
Change the setting after pre-run program is completed.
AOC control was attempted when the instrument and AOC were not connected.¶
Check the following items.
-
Connect the instrument and AOC.
-
AOC is installed correctly.
Installation_AOC-30i
Installation_AOC-20i -
AOC power is [On].
Calibration was performed in a situation where calibration was not possible.¶
Check the following items before calibration.
- Stopping GC
- Turn Off FLOW.
- Open each unit to the atmosphere, when the pressure higher than the specified value has already been applied.