[GC Stop Sequence] Screen
Press HOME button - [GC Start/Stop Sequence] while the GC is turned on to display [GC Stop Sequence] screen.
You can set parameters related to the GC shutdown such as stop time, flow off time, and sleep time.
| No. | Item | Description | ||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|   | Stop Time | Sets the period of time after  [Stop GC] is pressed until the instrument stops temperature/detector control. | ||||||||||||||||
| Range | 0.0 to 6000.0 min | |||||||||||||||||
| Default | 0.0 min | |||||||||||||||||
|   | Flow Control | Displays the current status of flow control. Press [Off] or [On] to switch the control status.
Select [On] to continue the carrier gas flow after the GC shutdown. Select [Off] to end it after the GC shutdown. Select [Off] to stop gas flow after the time set in  [Flow Off Time] elapses. | ||||||||||||||||
| Items | Off, On | |||||||||||||||||
| Default | On | |||||||||||||||||
|   | Flow Off Time | Sets the period of time between the end of temperature/detector control and the end of gas control.
This can be set when  [Flow Control] is [Off]. Note 
 | ||||||||||||||||
| Range | 0.0 to 6000.0 min | |||||||||||||||||
| Default | 0.0 min | |||||||||||||||||
|   | Sleep | When [Enable] is selected, the instrument will go into sleep mode automatically after the time set in  [Flow Off Time] elapses.
This can be set when  [Flow Control] is [Off]. | ||||||||||||||||
| Items | Enable, Disable | |||||||||||||||||
| Default | Disable | |||||||||||||||||
|   | Restart | The action when [Disable] or [Enable] is selected are different depending on the settings in  [Sleep]. 
  [Flow Control] is [Off]. | ||||||||||||||||
| Items | Enable, Disable | |||||||||||||||||
| Default | Disable | |||||||||||||||||
|   | Restart Time | Sets the time until the instrument will be automatically turned on.
Count starts after control of temperature and the detector is stopped. Count starts at the same timing as flow off time.
This can be set when  [Restart] is set at [Enable].
When the time is set at a value lower than  [Flow Off Time], the instrument restarts instead of going into sleep mode even if  [Sleep] is set at [Enable]. | ||||||||||||||||
| Range | 0.1 to 6000.0 min | |||||||||||||||||
| Default | 0.1 min | |||||||||||||||||
|   | Stop GC | When you press [Stop GC], the GC will stop according to [GC Stop Sequence] screen settings. If you press [Stop GC] while the instrument is performing analysis, GC stop is scheduled and [Abort] is displayed. Press [Abort] to cancel the GC stop. During the GC Shutdown, [GC Start Sequence] is displayed. Press [GC Start Sequence] to display [GC Start Sequence] screen. | ||||||||||||||||
Submenu¶
| Item | Description | 
|---|---|
| GC Start Sequence | Displays [GC Start Sequence] screen. | 
| File | Displays [File Select] screen. You can change the file to be used. | 
| Clean Up | Displays [Clean Up] screen. You can set parameters for clean up during the GC startup. | 
Examples for the GC stop¶
The flow off time should be changed depending on the column oven temperature.
The following examples show various GC stop sequence adapted to the conditions of the instrument.
To protect columns, set the time so that the carrier gas flow stops after each parts is cooled during the GC shutdown.
- Turn off heater switch immediately and stop the carrier gas after 20 minutes.
 Stop Time : 0 min
 Flow Control : Off
 Flow Off Time : 20 min
If the carrier gas stops while the column oven temperature is high, the liquid phase of the column may be degraded. The flow off time should be set at a value where the column oven temperature drops adequately.
- 
A column is conditioned (aging), and then the column oven is cooled. Carrier gas flow is then shut down. 
 Stop Time : Column conditioning time
 Flow Control : Off
 Flow Off Time : Approximately 20 min
- 
Turn off heater switch 10 minutes after [Stop GC] is pressed, keep the carrier gas flow, and restart the instrument after 20 hours (1200 minutes). 
 Stop Time : 10 min
 Flow Control : Off
 Flow Off Time : 1200 min (20 hours)
 Sleep : Disable
 Restart : Enable
 Restart Time : 1200 min (20 hours)
When the GC is turned off, the oven fan automatically stops based on settings in [Fan Off Temp].