{"title":"\u30a2\u30d7\u30ea\u30b1\u30fc\u30b7\u30e7\u30f3","columnTitle":"\u30a2\u30d7\u30ea\u30b1\u30fc\u30b7\u30e7\u30f3","source":"product","key":4890,"filter_types":["applications","application_note","posters"],"config_list":[],"anchor":"table_anchor_applications","filter":true} {"title":"\u30c6\u30af\u30cb\u30ab\u30eb\u30c9\u30ad\u30e5\u30e1\u30f3\u30c8","columnTitle":"\u30c6\u30af\u30cb\u30ab\u30eb\u30c9\u30ad\u30e5\u30e1\u30f3\u30c8","source":"product","key":4890,"filter_types":["technical","technical_reports","white_papers","primers"],"config_list":[],"anchor":"table_anchor_technical","filter":true} {"title":"\u5404\u7a2e\u8cc7\u6599","columnTitle":"\u5404\u7a2e\u8cc7\u6599","source":"product","key":4890,"filter_types":["manuals"],"config_list":[],"anchor":"table_anchor_manual","filter":true}

Easy wiper-Sを用いた有機物の回収試験

Easy wiper-Sを使用して「スワブサンプリング-直接燃焼炭素測定法」で各種有機物の回収率評価を行いましたのでご紹介します。

試料

清浄なステンレス板表面に各種有機物の溶液を塗り広げた後、溶媒を蒸発・乾燥させたものを模擬試料としました。(トラネキサム酸、無水カフェイン、イソプロピルアンチピリン、ニフェジピン、リンデロン軟膏、ワセリン)模擬試料の炭素量は0.200 mgとしました。

測定条件

分析計 全有機体炭素計 TOC-LCPH+固体試料燃焼装置 SSM-5000A(セル切り替えバルブセット付加システムでIC回路バイパス)
セル長 ショートセル
キャリアガス 400 mL/min 酸素ガス
測定項目 TC(全炭素)
検量線 1 %Cグルコース水溶液30 μL(0.300 mgC)による1点検量線
拭き取り方法 純水120~150 μLで湿らせたEasy wiper-Sをピンセットで操作して拭き取り

測定結果

試験した6種類の有機物すべてについて高い回収率を得らました。測定データの一部を図1に示します。

物質名 TC値/mg 回収率/ %
トラネキサム酸 0.207 104
無水カフェイン 0.203 102
イソプロピルアンチピリン 0.194 97
ニフェジピン 0.202 101
リンデロン軟膏 0.204 102
ワセリン 0.199 995
測定結果

図1 トラネキサム酸、無水カフェイン、ニフェジピンの測定結果