Inspection and Replacement of FID Jet

Required devices

No.
Parts name
Category
Required number
Note
Tweezers
1
Hexnut Driver For FID Nozzles
1
Options
FID Nozzle Φ0.3
1
φ0.3, standard
(for capillary column)
FID Nozzle Φ0.8
1
φ0.8, for analysis of aqueous solution
(for capillary column)
FID Nozzle ASSY for packed column
1
φ0.5, for packed column
Screwdriver
1
Stylet Wire (0.1mm×150mm)
1

Procedures

Step 1

On the instrument monitor or workstation, check that the temperature of the column oven is lower than 40 °C, and check that the temperatures of the injection port and detector are 50 °C or lower.
After checking the temperatures, turn OFF the main power to the instrument.

Step 2

Remove the column installed on FID.

Step 3

Open the INJ/DET cover (★1).

Step 4

Remove the stopper plate, and remove the igniter.

Step 5

Loosen the two screws, and remove the upper cover (★2), high voltage unit (★3), and collector (★4) together.

Hint

The screws have a fall-prevention structure, so removal is not necessary.

Step 6

Loosen the FID nozzle (③) with a hexnut driver (②).

Step 7

Remove the FID nozzle with tweezers (①).

Step 8

Inspect the area around the tip of the FID nozzle.

Hint

Soaking in organic solvent (hexane, acetone, etc.) is also effective.
Ultrasonic cleaning provides even better cleaning results.

Step 9

Install the FID nozzle(③) inside the FID body, tightening it securely with a hexnut driver (②).

Step 10

Attach the assembled collector, upper cover, and high voltage unit to the FID body, and tighten the two screws to fix them.

Note

Ensure that the screws are tightened alternately and gradually to apply even force to the collector.

Step 11

Attach the igniter, and fix it with the stopper plate.

Step 12

Close the INJ/DET cover.

Note

If the nozzle was cleaned using organic solvents, increase the detector temperature to 20–30°C above the normal operating temperature and perform a bake-out for about 3–4 hours prior to use.
Residual organic solvents can contribute to noise and other issues.