Installation of Capillary Column (Graphite Ferrule)

Video (Installation of Graphite Ferrule to Column)

Video (Installation of Column to GC)

Required devices

No.
Parts name
Category
Required number
Note
①
10x12 Spanner
1
②
6x8 Spanner
2
③
Tweezers
1
④
Cutter, Capillary Tube
1
⑤
Non Slit Nut
1
⑥
Slit Nut
1
⑦
Graphite Ferrule
2
⑧
Graphite jig (INJ)
1
⑨
Graphite jig (DET)
1
Note

A graphite ferrule that matches the inner diameter of the column must be selected.
Refer to Ferrule Type .
Note that this procedure cannot be used for GCMS.

Procedures

Hint

When performing Direct analysis, refer to Column Installation for Direct Analysis

Step 1

Stop the instrument. Then, on the instrument monitor or workstation, check that the temperature of the column oven is lower than 40 °C, check that the temperatures of the injection port and detector are 50 °C or lower, and check that the gas supply is stopped and there is no residual pressure.

Step 2

Slide the injection port side of the column through the column nut (without slit) (⑤), and then through the graphite ferrule (⑦).

Note

Organic matter easily adheres to graphite. If you directly touch graphte ferrule, it may affect the analytical data.

Step 3

Fix the graphite ferrule temporarily with the Graphite jig (⑧) and a 6×8 spanner (②).

In this case, push out capillary column about 10 mm from the Graphite jig.

Note
  • For SPL/MMI/WBI, the insertion length when using a graphite ferrule is 34 mm.

  • When using Graphite jig for SPL/MMI/WBI, the sealed end of the graphite ferrule is fixed at 34 mm from the column end.

  • When performing splitless analysis using a wide bore column with an inner diameter of 0.53 mm, use the Graphite jig and WBI so that the sealed end of the graphite ferrule is fastened at 15 mm from the column end.

Step 4

Cut the end of column with capillary cutter (④) etc.

Caution

Cut the column so that the end becomes flat.

When the graphite excess into the graphite adjuster, remove the entered graphite with tweezers, etc.
Caution should be exercised not to damage the column.

If the column is installed while excess graphite is still attached onto the graphite adjuster is not removed, the graphite may remain and cause clogging in the capillary adapter after you remove the column. When the graphite is clogged, disconnect the capillary adopter and remove the clogging with thin cotton swabs, etc.

Step 5

Prepare the detector side of the column in the same way.

Normally, a nut with slit is used on the detector side, therefore, fix only the graphite ferrule with the Graphite jig (⑨). (If analyses are performed with column temperature set at more than 350 °C, thermal capacity may need to be reduced by using a small nut without slit, which is the same as the injection port side.)

Note

The position where the graphite ferrule is fixed depends on the detector to be used.
Fix the graphite ferrule with the adjuster for each detector.

GC-2060

Detector
Position
Note
FID
54 mm
FID (when using Jetanizer)
40 mm
STCD
68 mm
FPD
82 mm
FTD
69 mm
Shorten by 2–3 mm when using a metal capillary column.
BID
74 mm
ECD
32 mm
Graphite ferrule jig, ECD_32mm P/N 221-48610-04 is used.

*Method for distinguishing the different ECD flow paths
ECD-2010 Exceed column insertion length differs depending on the flow path.
Refer to ECD-2010 Exceed Method for distinguishing the different ECD flow paths to determine the flow path.

Step 6

Pull the column oven door latch toward you to open the column oven door.

Step 7

Hang the column on the column hanger.

Step 8

Insert the column into the injection inlet.

Step 9

Turn the nut counterclockwise by hand to tighten it.

Step 10

Tighten it about 1/2 turn with a 6×8 spanner (②).

Step 11

Insert the column also on the detector side.

Step 12

Thread a nut (with slit) (⑥), and turn the nut clockwise by hand to tighten it.

Step 13

Tighten it about 1/2 turn with a 10×12 spanner (①).

Step 14

Close the column oven door.

Step 15

Start the GC.