Column Replacement Procedure (HS-20 L.T)

Required devices

No.
Parts name
Category
Required number
Note
①
Wrench (1/4 inch x 5/16 inch)
1
②
Capillary cutter
1
③
Screw driver
1
④
Leak detector
1
⑤
Soft cloth
1
⑥
Organic solvent (Toluene, Methanol, etc.)
1
⑦
Loupe
1
⑧
Ruler
1
⑨
Septum
1
⑩
Capillary column
1
Contact SHIMADZU GLC.
⑪
Transfer line capillary tube
1
⑫
Press-Tight
1
⑬
Vespel ferrule
1
⑭
Nut, SSNE16/012
1

Connecting a column to the HS-20 L.T

Connect the HS-20 L.T and column by using a Press-Tight connector across a retention gap as shown below.

Procedures

Step 1

When using the HSS control software, double-click the [HS-20] icon to start the HSS control software.

Hint

The HSS control software opens with a double-click on “HSS20.exe ” in the following folder:
C:\Program Files (x86)\Shimadzu\HS20

Step 2

The [Select Instrument] window is displayed. Click the [OK] button.

Step 3

Ensure the temperature of each part of HS-20 (e.g. oven temperature, sample line temperature,transfer line temperature) is reduced to room temperature.

Wear safety goggles.

Step 4

For GCMS, perform auto shutdown using the workstation. Wait until the "Completed" message appears.

For GC, stop the instrument. Then, on the instrument monitor or workstation, check that the temperature of the column oven is lower than 40 °C, check that the temperatures of the injection port and detector are 50 °C or lower, and check that the gas supply is stopped and there is no residual pressure.

Step 5

Once the system has cooled down, close the workstation and HSS control software and turn off the power switch on HS-20.

The main power must be turned off and the temperature of each instrument part must be at room temperature before the cover of HS-20 can be opened.

Step 6

Pull the lever at the lower right of the GC front door to open the oven door.

Step 7

Cut the transfer line capillary tube and the capillary column prior to their installation into the Press-Tight.

Step 8

Remove the fastening screw M4 (③) and the knurled screw (④) on the maintenance hatch (①) side panel.
Loosen the knurled screw (②) on the maintenance hatch (①) rear panel.

Step 9

Slide the maintenance hatch forward and lift it to disassemble.

Step 10

Remove the 2 locks of the valve cover and take off the cover.

Step 11

Remove the aluminum inner cover (★1).
The aluminum inner cover (★1) is simply placed on the valve and is not fastened.

Step 12

The valve tube area is exposed.
Sample loop (★2), switching valve (★3), split joint (★4)

Step 13

Use a wrench (1/4 inch × 5/16 inch) to remove the split joint nut SSNE16/012 (★5).

Step 14

Prepare a new transfer line capillary tube that matches the inside diameter of the capillary column.

Step 15

Cut both ends of the transfer line capillary tube at a right angle (i.e. orthogonal) using a capillary cutter and make it to about 2 mm.
Cut the the capillary column (i.e. the end that goes into the Press-Tight) at a right angle (i.e. orthogonal).

Note

Check that the column is scored correctly using a loupe.

Step 16

Before attaching the transfer line capillary tube, thoroughly wipe any dirt off the tip of the transfer line capillary tube with a piece of gauze moistened with organic solvent (e.g. methanol).

Step 17

Connect the new transfer line capillary tube with the nut and the ferrule.
The end of the column should protrude 35 to 40 mm from the ferrule.

Step 18

Install the aluminum inner cover (★1).
The aluminum inner cover (★1) is simply placed on the valve and is not fastened.

Step 19

Secure the two locks (★7) of the valve cover (★6).

Step 20

Place the maintenance hatch and slide it forwards until the rear side groove fits into the screw. Once the groove is fit, slide it to the back.

Step 21

Tighten the knurled screw (②) on the maintenance hatch (①) rear panel.
Install the fastening screw M4 (③) and the knurled screw (④) on the maintenance hatch (①) side panel.

Step 22

Insert the tip of the transfer line capillary tube until it is firmly pressed in the Press-Tightconnector.
Apply a moderate amount of force when inserting.
If too much pressure is applied, the tip of the capillary column may be crushed.

Step 23

Insert the end of the capillary column into the other side of the Press-Tight.


Step 24

Attach the capillary column so that it does not touch the inner wall of the column oven.

Step 25

If you changed a capillary column, register or change the capillary column in workstation software.

Step 26

For GCMS, perform auto startup using the workstation.

Step 27

For GCMS, monitor the water and nitrogen in the peak monitor window, and check for leaks.

Note

If there is a leak, it is recommended to use a leak detector to check for leakage in the Press-Tight connector.

Step 28

Check for leakage in the Press-Tight connector by using a leak detector.

Note

Do not set the column oven temperature higher than the upper limit of the operating temperature of the Press-Tight.