Adjusting a Crimper
Required devices¶
No. |
Parts name |
Category |
Required number |
Note |
|---|---|---|---|---|
① |
Hand crimper (Electronic crimper) |
1 |
||
② |
Gloves (thick) |
1 |
||
③ |
Hex wrench |
1 |
||
④ |
Vial |
1 |
||
⑤ |
Vial cap |
1 |
Procedures¶
Since a high-temperature cap will be exposed to a high temperature, the cap may leak if tightened in the same way as the pressure-release type.
A high-temperature cap must be sealed more tightly than the pressure-release type.
The crimper must be adjusted according to the type of vial cap (e.g. pressure-release cap, high-temperature cap).
Step 1¶
Before trying on samples, check the crimper operation first with an empty vial according to
Crimping a Vial Cap
.
If there is no problem found with the crimper operation, the following crimper adjustment steps can be omitted.
Step 2-1¶
When using a manual crimper such as the type shown below
-
If the crimp is too weak, use a hex wrench to adjust the crimper.
Turn the adjustment screw to the left as shown in the bottom image on the left.
-
If the crimp is too strong, use a hex wrench to adjust the crimper.
Turn the adjustment screw to the right.
When changing from the pressure-release vial cap to the high-temperature vial cap
Using a hex wrench, turn the crimper adjustment screw once to the left from the position for the pressurerelease vial cap.
|
|
|---|---|
|
Do not turn the adjustment screw too far to the left. |
Step 2-2¶
When using a manual crimper such as the type shown below
-
If the crimp is too weak, turn the orange adjustment knob in the + direction.
-
If the crimp is too strong, turn the orange adjustment knob in the - direction.
|
|
|---|---|
|
Do not turn the adjustment knob too far in the + direction. |
Step 2-3¶
When using an electronic crimper such as the type shown below
- If the crimp is too weak, press the + button.
If you press the + button, the Status LED will light amber. - If the crimp is too strong, press the - button.
If you press the - button, the Status LED will light green.
|
|
|---|---|
|
Do not press the + button too many times. |
|
Do not use an electronic crimper or decapper with a high-temperature vial cap as the vial may become damaged. |
Step 2-4¶
When using an electronic crimper such as the type shown below
- If the crimp is too weak, press the ▽ button.
If you press the ▽ button, the percentage of the LCD panel rises. - If the crimp is too strong, press the △ button.
If you press the △ button, the percentage of the LCD panel decrease.
|
|
|---|---|
|
Do not press the ▽ button too many times. |
|
Do not use an electronic crimper or decapper with a high-temperature vial cap as the vial may become damaged. |
Step 3¶
Once a crimper and decapper are adjusted, sealing should be checked with an empty vial before trying on
samples according to
Crimping a Vial Cap
.
Repeat the adjustments until a proper sealing is obtained.