AOC-20s expansion and control error [Error Code:6015]
Cause¶
The following causes are possible.
- The extension motor has a problem.
- A vial with dimensions different from those of the vial used for teaching is used.
Solution¶
Redo the teaching using the vial to be used for actual analysis.
In case of AOC-30i¶
AOC-30i_AOC-20sU_SamplerTeaching