DET#n set temperature is low [Error Code:1049 1050 1051 1052]
Cause¶
Detector temperature is set at a lower value than column temperature.
Solution¶
- To prevent detector contamination, set a detector temperature higher than the column oven temperature.
- When setting a detector temperature lower than the column oven temperature, turn off the contamination prevention function.
- [HOME] - [Function] - [Configuration] - [Max Temperature] to display [Max Temperature] screen.
- Display the second page and turn off the contamination prevention function of the relevant detector.