{"title":"\u30a2\u30d7\u30ea\u30b1\u30fc\u30b7\u30e7\u30f3","columnTitle":"\u30a2\u30d7\u30ea\u30b1\u30fc\u30b7\u30e7\u30f3","source":"product","key":3224,"filter_types":["applications","application_note","posters"],"config_list":[],"anchor":"table_anchor_applications","filter":true} {"title":"\u30c6\u30af\u30cb\u30ab\u30eb\u30c9\u30ad\u30e5\u30e1\u30f3\u30c8","columnTitle":"\u30c6\u30af\u30cb\u30ab\u30eb\u30c9\u30ad\u30e5\u30e1\u30f3\u30c8","source":"product","key":3224,"filter_types":["technical","technical_reports","white_papers","primers"],"config_list":[],"anchor":"table_anchor_technical","filter":true} {"title":"\u5404\u7a2e\u8cc7\u6599","columnTitle":"\u5404\u7a2e\u8cc7\u6599","source":"product","key":3224,"filter_types":["manuals"],"config_list":[],"anchor":"table_anchor_manual","filter":true}

LPG中の微量水分測定

島津が提案しているWK59649測定法はガスクロマトグラフィー(GC)およびBIDを用いたLPG中の微量水分測定の標準試験法であり,米国の公定法として申請中です。従来のTCDを用いたシステムよりも100倍程度の感度を有しています。 このメソッドではNexis GC-2030,BID,ガスサンプリングバルブ,液体サンプリングバルブ,Watercolカラムを使用しています。

LPG中水分析クロマトグラム(25ppm)

LPG中微量水分測定クロマトグラム(25ppm)
定量限界(S/N = 10)および検出限界(S/N = 3.3)は,それぞれ0.66ppmおよび0.22ppm

本システムにおける微量水分測定(25ppm)の再現性は,表1に示す通りRSD2%未満でした。 また,サンプルにエチルメルカプタンのような硫黄成分が含まれていても,水分を分離することができ,正確な定量を行うことができます。

表1 微量水分測定再現性(n=5)

Water No.1 No.2 No.3 No.4 No.5 Average Standard
Deviation
%RSD
RT(min)  14.285 14.288 14.286 14.296 14.304 14.292 0.008 0.06
Area(µV・s)  244,037  249,854 246,884 242,950 238,428 244,430 4,296 1.76
 Height(µV)  12,418 12,600 12,468 12,045 11,851 12,276 315 2.57
Conc.(ppm)  24.96 25.56 25.25 24.85 24.39 25.00 0.44  1.76
WatercolはSigma-Aldrich社の商標です。