{"hasMore":true,"more":{"id":"tbaleAnchor_more","label":"\u7d9a\u304d"},"anchors":[{"id":"tbaleAnchor_applications","label":"\u30a2\u30d7\u30ea\u30b1\u30fc\u30b7\u30e7\u30f3"},{"id":"tbaleAnchor_technical","label":"\u30c6\u30af\u30cb\u30ab\u30eb\u30c9\u30ad\u30e5\u30e1\u30f3\u30c8"},{"id":"tbaleAnchor_manual","label":"\u30de\u30cb\u30e5\u30a2\u30eb"}]} {"title":"\u30a2\u30d7\u30ea\u30b1\u30fc\u30b7\u30e7\u30f3","columnTitle":"\u30a2\u30d7\u30ea\u30b1\u30fc\u30b7\u30e7\u30f3","source":"product","key":4803,"filter_types":["applications","application_note","posters"],"config_list":[],"anchor":"tbaleAnchor_applications","filter":true} {"title":"\u30c6\u30af\u30cb\u30ab\u30eb\u30c9\u30ad\u30e5\u30e1\u30f3\u30c8","columnTitle":"\u30c6\u30af\u30cb\u30ab\u30eb\u30c9\u30ad\u30e5\u30e1\u30f3\u30c8","source":"product","key":4803,"filter_types":["technical","technical_reports","white_papers","primers"],"config_list":[],"anchor":"tbaleAnchor_technical","filter":true} {"title":"\u30de\u30cb\u30e5\u30a2\u30eb","columnTitle":"\u30de\u30cb\u30e5\u30a2\u30eb","source":"product","key":4803,"filter_types":["manuals"],"config_list":[],"anchor":"tbaleAnchor_manual","filter":true}

アプリケーション

灯軽油中のクマリン分析

灯軽油中のクマリンを分析しました。
灯軽油は夾雑成分が多く影響を受けやすい試料ですが,ハートカット分析により試料の前処理無しで目的の成分を完全分離しました。

「不正軽油(灯油50%軽油50%)」10回連続分析結果

注入量:1µL
Index クマリン
面積値 (uv.s)
灯油50%混合軽油 1 1,364
灯油50%混合軽油 2 1,389
灯油50%混合軽油 3 1,374
灯油50%混合軽油 4 1,372
灯油50%混合軽油 5 1,369
灯油50%混合軽油 6 1,381
灯油50%混合軽油 7 1,363
灯油50%混合軽油 8 1,369
灯油50%混合軽油 9 1,369
灯油50%混合軽油 10 1,367
%RSD 0.580522
最大 1,389
最小 1,363

マルチカット性能

Grob Test Mix(溶媒 + 8成分)を用いて複数回ハートカットが1stクロマトグラムの保持時間に与える影響を評価しました。複数回のハートカットを行っても後半のクロマトグラムに与える影響が僅かであることを確認しました。