{"hasMore":true,"more":{"id":"tbaleAnchor_more","label":"\u7d9a\u304d"},"anchors":[{"id":"tbaleAnchor_applications","label":"\u30a2\u30d7\u30ea\u30b1\u30fc\u30b7\u30e7\u30f3"},{"id":"tbaleAnchor_technical","label":"\u30c6\u30af\u30cb\u30ab\u30eb\u30c9\u30ad\u30e5\u30e1\u30f3\u30c8"},{"id":"tbaleAnchor_manual","label":"\u30de\u30cb\u30e5\u30a2\u30eb"}]} {"title":"\u30a2\u30d7\u30ea\u30b1\u30fc\u30b7\u30e7\u30f3","columnTitle":"\u30a2\u30d7\u30ea\u30b1\u30fc\u30b7\u30e7\u30f3","source":"product","key":3280,"filter_types":["applications","application_note","posters"],"config_list":[],"anchor":"tbaleAnchor_applications","filter":true} {"title":"\u30de\u30cb\u30e5\u30a2\u30eb","columnTitle":"\u30de\u30cb\u30e5\u30a2\u30eb","source":"product","key":3280,"filter_types":["manuals"],"config_list":[],"anchor":"tbaleAnchor_manual","filter":true}

コールタールの網羅的分析および硫黄化合物の選択的検出

GCxGCはその分離能の高さから,試料の網羅的測定に利用されます。 しかし,極めて複雑な試料では不十分な場合もあります。 例えば,ジクロロメタンで希釈したコールタールを直接GCxGC-MS/MS(Q3 Scan)で測定すると,非常に複雑な2Dマップとなります(Fig.1)。 Fig.2 ~ 4は,同じ試料をLC-GCxGC-MS/MSで測定した結果です。 LCでタイプ毎に分離された3画分を連続してそれぞれGCxGC-MS/MS測定しました。 HPLCによる分離で簡素化された2Dマップからより詳細な網羅的分析が可能となります。

また,化石燃料中の硫黄化合物は,種類も豊富で低濃度で存在するため,マトリクスの影響を受けやすいことから,高選択・高感度検出が威力を発揮します。
Fig.5は,2画分目に含まれるS化合物(Dibenzothiophene)をMRM測定した結果です。MS/MSによりジベンゾチオフェンの各異性体を選択的に検出できました。

Fig.5 2画分目のDibenzothiopheneの2Dマップ(MRM)

 

* データは,Mondello先生(Messina大学,Chromareleont S.r.l.)からご提供いただきました。