Range of Linear Temperature Rise

100 V (20 A mode)

Range of temperature Rate of temperature rise
to 70 ℃ 0 ℃ to 65 ℃/min
to 115 ℃ 0 ℃ to 50 ℃/min
to 200 ℃ 0 ℃ to 40 ℃/min
to 250 ℃ 0 ℃ to 25 ℃/min
to 300 ℃ 0 ℃ to 20 ℃/min
to 380 ℃ 0 ℃ to 15 ℃/min
to 450 ℃ 0 ℃ to 7 ℃/min

100 V (30 A mode), 115 V

Range of temperature Rate of temperature rise
to 70 ℃ 0 ℃ to 65 ℃/min
to 115 ℃ 0 ℃ to 50 ℃/min
to 200 ℃ 0 ℃ to 40 ℃/min
to 300 ℃ 0 ℃ to 30 ℃/min
to 380 ℃ 0 ℃ to 15 ℃/min
to 450 ℃ 0 ℃ to 10 ℃/min

200 V

Range of temperature Rate of temperature rise
to 70 ℃ 0 ℃ to 85 ℃/min
to 115 ℃ 0 ℃ to 65 ℃/min
to 175 ℃ 0 ℃ to 50 ℃/min
to 250 ℃ 0 ℃ to 45 ℃/min
to 300 ℃ 0 ℃ to 35 ℃/min
to 380 ℃ 0 ℃ to 25 ℃/min
to 450 ℃ 0 ℃ to 20 ℃/min

230 V

Range of temperature Rate of temperature rise
to 70 ℃ 0 ℃ to 120 ℃/min
to 115 ℃ 0 ℃ to 95 ℃/min
to 175 ℃ 0 ℃ to 80 ℃/min
to 300 ℃ 0 ℃ to 55 ℃/min
to 450 ℃ 0 ℃ to 35 ℃/min