CARx insert counter error [Error Code:84 85 86 87]

Cause

The number of analysis exceeded the set threshold.

Solution

Replace the consumables and then reset the counter on [Analysis Counter] screen.

Refer the below for how to replace the liner.

Replacement of Glass Insert

Refer to the following for how to reset the counter value.

When executing from the GC main unit screen

Press HOME button - [Function] - [Diagnosis] - [Analysis Counter] to display [Analysis Counter] screen.
Press [Reset] to reset the count.

When executing from the LabSolutions

Press [System Check] - [Reset Consumables] on the assistant bar, and use the Liner Usage Counter [Reset] button to reset.