Discharge gas flow rate is not enough [Error Code:4213]
Cause¶
The pressure or flow rate may have not reached the set value due to the following reasons.
- The gas supply pressure is insufficient.
- There is a leak at the pipe connection.
- The resistance of the supply piping is high (the connected piping is long or the inner diameter of the piping is narrow).
Solution¶
- Check that the gas main valve is open.
- Check that the gas supply pressure is within the range below.
- When using the helium purifier HP2, check that there is no leak at tubing connections.
- Check for any looseness or leaks at the pipe connections.
- Contact your Shimadzu service representative to check the inner diameter and length of used piping.