DET#n FTD overcurrent detected [Error Code:4105 4106 4107 4108]

Cause

Detector current is abnormal.
If the resistance of the filament becomes abnormally high when the FTD is in use, the protection circuit is actuated to prevent damage to the filament.

Solution

Turn off the main power of the instrument, and confirm the following items.

Remove the cause of the error, then turn on the instrument.
If the error cannot be resolved, turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative.