xxx temperature exceeds the limit [Error Code:1002 1003 1004 1005 1006 1007 1008 1009 1053 1062 1065]
Cause¶
Temperature exceeds the maximum allowable value.
Solution¶
Please follow the steps below;
-
Press [Ignore] to close the error dialog.
-
Press HOME button - [Function] - [Configuration] - [Max Temperature] to display [Max Temperature] screen and change the maximum temperature value.
[Max Temperature] Screen -
After changing the value, press [Reset] in the error dialog.