Entering Maintenance Mode (with Monitor)

For models equipped with a monitor display, such as the GC-2030, the instrument can be switched to maintenance mode via the monitor. (INJ Maintenance function)

Required devices

Procedures (Entering Maintenance Mode)

Step 1

Press

(HOME) - [Monitor] - [Maintenance] and [Start].

 

When [Start] is pressed, injection port temperature decreases automatically. When injection port temperature decreases adequately and the instrument is ready for maintenance, the message [GC is ready for maintenance] appears and [Start] changes to [End].

Procedures (Exiting Maintenance Mode)

Step 1

Press [Reset] of [Septum Counter] on GC touch panel.

Step 2

Press [End].

Automatically supplies carrier gas, and starts temperature control of the inlet and the column oven.