Controllable ranges of temperature program rates
100 V(15 A mode)
| Temperature |
Control range |
| to 70 ℃ |
0 ℃ to 50 ℃/min |
| to 115 ℃ |
0 ℃ to 30 ℃/min |
| to 175 ℃ |
0 ℃ to 15 ℃/min |
| to 200 ℃ |
0 ℃ to 10 ℃/min |
| to 300 ℃ |
0 ℃ to 7 ℃/min |
| to 350 ℃ |
0 ℃ to 4 ℃/min |
100 V(20 A mode), 115 V
| Temperature |
Control range |
| to 70 ℃ |
0 ℃ to 80 ℃/min |
| to 115 ℃ |
0 ℃ to 50 ℃/min |
| to 200 ℃ |
0 ℃ to 40 ℃/min |
| to 250 ℃ |
0 ℃ to 25 ℃/min |
| to 300 ℃ |
0 ℃ to 20 ℃/min |
| to 380 ℃ |
0 ℃ to 15 ℃/min |
| to 450 ℃ |
0 ℃ to 7 ℃/min |
200 V
| Temperature |
Control range |
| to 70 ℃ |
0 ℃ to 85 ℃/min |
| to 115 ℃ |
0 ℃ to 65 ℃/min |
| to 175 ℃ |
0 ℃ to 50 ℃/min |
| to 250 ℃ |
0 ℃ to 40 ℃/min |
| to 300 ℃ |
0 ℃ to 30 ℃/min |
| to 380 ℃ |
0 ℃ to 25 ℃/min |
| to 450 ℃ |
0 ℃ to 20 ℃/min |
230 V
| Temperature |
Control range |
| to 70 ℃ |
0 ℃ to 120 ℃/min |
| to 115 ℃ |
0 ℃ to 95 ℃/min |
| to 175 ℃ |
0 ℃ to 65 ℃/min |
| to 300 ℃ |
0 ℃ to 45 ℃/min |
| to 450 ℃ |
0 ℃ to 35 ℃/min |