Detector Gas Flow Rate

Set flow rate of detector gas using values shown in the following table as a reference.

Hydrogen
Air
Makeup Gas
(BID: DCG)
(ECD: ECD gas)
Note
FID
(Capillary column)
32.0 mL/min
200.0 mL/min
24.0 mL/min
-
FID
(Packed column)
32.0 mL/min
200.0 mL/min
Total flow rate of carrier gas and makeup gas:
20 to 50 mL/min
-
FTD
(N-mode)
1.0 to 2.0 mL/min
140.0 to 150.0 mL/min
10.0 to 30.0 mL/min
-
FTD
(P-mode)
3.0 to 6.0 mL/min
140.0 to 150.0 mL/min
10.0 to 30.0 mL/min
-
FPD
(P-mode)
62.5 mL/min
90.0 mL/min
-
-
FPD
(S-mode)
40.0 mL/min
60.0 mL/min
-
-
FPD
(Sn-mode)
75.0 mL/min
65.0 mL/min
-
-
ECD
-
-
15.0 mL/min
Nitrogen (N2) is used
BID
-
-
Depends on the column flow rate.
Less than 10.0 mL/min:50.0 mL/min
10.0 to 15.0 mL/min:70.0 mL/min
More than 15.0 mL/min:100.0 mL/min
Helium (He) is used
SCD
80.0 mL/min
-
40.0 mL/min
Oxygen for the furnace 10.0 mL/min
Oxygen for the ozone generator 25.0 mL/min