The config is such that auto eco cannot be executed [Error Code:9086]

Cause

Auto Eco was executed in one of the following cases:

Solution

If the PREP RUN function is [Disable], follow the steps below to [Enable] it.
Press the HOME button > [Function] > [PREP RUN] to display the [PREP RUN] screen and change the PREP RUN settings.

If the PREP RUN function is enabled but still cannot be executed, it is connected to the detector that does not support Auto Eco operation (e.g. QP2020NX, TQ8040NX, TQ8050NX).