AOCm syringe counter warning [Error Code:9059 9060]

Cause

Exceeded the set threshold.

Solution

Replace the syringe and then reset the counter.

Reset the counter

GC instrument monitor setting

Press HOME button - [Function] - [Diagnosis] - [Analysis counter], and then display [Analysis counter] screen and reset the syringe usage counter.

LabSolutions setting

Press [System Check] - [Consumables Reset] on the assistant bar, and reset the counter from the [Reset] button.