COL CRG use time warning [Error Code:9000]

Cause

Exceeded the set threshold of CRG coolant consumption (Operating Time).

Solution

Replace the coolant and then reset the counter.

GC instrument monitor setting

Press HOME button - [Function] - [Diagnosis] - [Coolant Consumption], and then display [Coolant Consumption] screen and reset the consumption (min).

LabSolutions setting

Press [System Check] - [Reset Consumables] on the assistant bar, and reset the consumption time (min) from the [Reset] button.

How to set the threshold for the coolant consumption time

The coolant consumption time depends on the gas cylinder volume and the temperature settings. Set the threshold according to the analysis conditions.
Perform analysis and check the actual coolant consumption before you set the threshold.