AOCn syringe counter error [Error Code:88 89]

Cause

The number of analysis exceeded the set threshold.

Solution

Replace the consumables and then reset the counter on [Analysis Counter] screen.

Refer to the following for how to reset the counter value.

When executing from the GC main unit screen

Press HOME button - [Function] - [Diagnosis] - [Analysis Counter] to display [Analysis Counter] screen.
Press [Reset] to reset the count.

When executing from the LabSolutions

Press [System Check] - [Reset Consumables] on the assistant bar, and use the Syringe Usage Counter [Reset] button to reset.