AOCn syringe counter error [Error Code:88 89]
Cause¶
The number of analysis exceeded the set threshold.
Solution¶
Replace the consumables and then reset the counter on [Analysis Counter] screen.
Refer to the following for how to reset the counter value.¶
When executing from the GC main unit screen¶
Press HOME button - [Function] - [Diagnosis] - [Analysis Counter] to display [Analysis Counter] screen.
Press [Reset] to reset the count.
When executing from the LabSolutions¶
Press [System Check] - [Reset Consumables] on the assistant bar, and use the Syringe Usage Counter [Reset] button to reset.