AOC-20s vial is not set [Error Code:6022]
Cause¶
There was no sample vial at the specified vial number position.
The following causes are possible.
- The specified vial No. was a position on the tray with no vial.
- The sample vials on the turret have been removed during analysis.
- Teaching position setting and the actual placement do not match (if used with a GC other than GC-2030).
Solution¶
Check and perform the following.
- Check the vial No. setting and the actual vial placement.
- Check the teaching position setting (if used with a GC other than GC-2030).