AOCm plunger error [Error Code:6003 6025]
Cause¶
The following causes are possible.
- Plunger is bent.
- Plunger is stuck.
- The plunger motor or sensor has a problem.
Solution¶
- If the plunger is bent, replace the syringe.
- If the plunger is not working properly, clean the plunger or replace the syringe.
- Revise the wash count and/or wash method as necessary.
- If the sample remains inside the syringe: Increase the number of washes. Increase the types of solvents used for cleaning.
- If the plunger is worn: Reduce the number of cleanings.
- If no abnormality is found in the plunger, and/or if the error occurs again after taking the measures, turn off the main power of the device and contact our sales office, distributor, or service representative designated by us.