AOCm rack error [Error Code:6001 6023]

Cause

The turret motor or sensor has a problem.

Solution

In case of AOC-30i

Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative.

In case of AOC-20i

After removing the turret and cleaning the rack gear, set the turret again and press the RESET key.
If the error appears even after checking the above, turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative.