Analysis waiting [Error Code:5112]
Cause¶
The analysis was started while the PREP RUN function is active and the instrument is in the waiting status.
Solution¶
Set [Prep Run Wait Time] longer.
The PREP RUN is activated when using the following functions.
- Splitless (split in standby)
- Carrier gas saver (saved in standby)
- CRG (coolant blowing stops in standby)
- High pressure injection (normal pressure in standby)
- Back flush (back flush in standby)
The instrument will go into a standby state if no analysis starts when the [Prep Run Wait Time] elapses after the analysis preparation starts, due to extended pre-run time, etc. It is therefore necessary to set [Prep Run Wait Time] longer than pre-run time.
For headspace analysis, set [Prep Run Wait Time] longer than heat-retention time.