CARm calc. ratio out of range [Error Code:5075 5076 5108 5109]

Cause

The carrier gas total flow calculated from split ratio program, etc. is outside the set range.

Solution

Change the split ratio so that the following equation is satisfied.
(Total flow) = (Column flow) + (Column flow * Split ratio) + (Purge flow)