DET#n STCD FW is unwitten[Error Code:4163 4164 4165 4166]

Cause

An error has occurred in the STCD-CPU.

Solution

Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative to ask for part replacement.