DetAPCn makeup valve not completely closed [Error Code:2057 2060 2063 2066]
Cause¶
There is a leak inside the makeup gas control valve of the APC for the detector.
Solution¶
Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative.