xxx MMI has been cooled for a long time [Error Code:1076 1077]
Cause¶
Cooling was stopped to protect the hardware because it was continuously cooled at temperatures below -50°C for an extended period.
| Temperature | Timeout (Approx.) |
|---|---|
| -50°C | 30 minutes |
| -100°C | 5 minutes |
| -160°C | 2 minutes |
Solution¶
You can avoid prolonged cooling in low-temperature ranges by shortening the pre-run time or equilibration time, or by increasing the heating rate of the temperature program in the injection port.