xxx sensor use time warning [Error Code:9012 9013 9014 9015 9016 9017 9018 9019 9024]

Cause

Risk of current leakage, heat, smoke, and ignition increases because usage time exceeds the set value.

Solution

Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative to ask for part replacement.