CARx septum counter warning [Error Code:9004 9005 9006 9052]

Cause

Exceeded the set threshold of inlet septum counter.

Solution

Replace the septum and then reset the counter.

Septum replacement

Replacement of Septum

Counter reset

GC instrument monitor setting

Press HOME button - [Function] - [Diagnosis] - [Analysis counter], and then display [Analysis counter] screen and reset the liner usage counter.

LabSolutions setting

Press [System Check] - [Consumables Reset] on the assistant bar, and reset the counter from the [Reset] button.