CARx calc. flow out of range [Error Code:5009 5010 5011 5073 5074 5077 5078 5102 5106 5107 5110 5111]
Cause¶
The total flow calculated from the inputted split ratio, column inlet pressure, column flow, linear velocity, and purge flow is outside the set range.
Solution¶
Set the split ratio, column flow, and purge flow so that the total flow is within the AFC control range.
Total flow is calculated with the following equation:
(Total flow) = (Column flow) + (Column flow * Split ratio) + (Purge flow)
TechnicalDocument_AFCnAPC-Control-Ranges-TotalFlow-n-InletPressure