Clean up program start error [Error Code:4305]
Cause¶
Clean up program cannot be started because oven temperature is not ready.
Solution¶
Set the initial temperature of the column oven at higher values than room temperature by more than 2 °C.
(Example, Set the temperature at more than 27 °C when the room temperature is 25 °C)
If the error appears repeatedly even after checking the above, turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative.