CARn was overpressure [Error Code:2214 2215 2216 2217]

Cause

Excessively high pressure is applied to the injection port. These errors may occur when supply gas pressure is changed during AFC operation.

Solution

If the error appears repeatedly even after checking the above, turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative.