CARm is not controlled [Error Code:2141 2142 2143 2213]
Cause¶
Temperature control started when the carrier gas control was turned off.
Solution¶
Turn on the carrier gas control or remove unused carrier gas line from the analytical line.
How to set Flow control [On/OFF]¶
Press HOME button - [Injection Port], and display below screen. Set turn "Flow" on ❶. When using two or more analytical lines, change analytical line.
How to change analytical line¶
Press HOME button - [Settings]- [Submenu]- [LINE Configuration] to display [LINE Configuration] screen.
Remove check for unused units.