Intake fan error [Error Code:102]

Cause

Dust has accumulated on the air inlet fan, or the fan may have failed.

Solution

Turn off the main power of the instrument, check for any foreign matter obstructing the air inlet fan operation, and remove it if possible.
If the problem persists, turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative to ask for part replacement.