Power frequency error [Error Code:101]

Cause

There is too much noise in the installation environment, or PCB may have failed.

Solution

Check for noise sources in the power supply and surrounding environment, and eliminate the if possible.
If the problem persists, turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative to ask for part replacement.