以下是检测器气体流速的标准值.
|
氢气 |
空气 |
尾吹气 |
|
FID |
40.0 mL/min |
400.0mL/min |
30.0mL/min |
|
FTD |
1.0 - 2.0 mL/min |
140.0 - 150.0 mL/min |
10.0 - 30.0 mL/min |
|
FTD |
3.0 - 6.0 mL/min |
14.00 - 150 mL/min |
10.0 - 30.0 mL/min |
|
FPD |
60.0 mL/min |
70.0 mL/min |
-- |
|
FPD |
80.0 mL/min |
110.0 mL/min |
-- |
柱流速超过 |
FPD |
80.0 mL/min |
120.0 mL/min |
-- |
|
FPD |
85.0 mL/min |
110.0 mL/min |
-- |
柱流速 |
FPD |
100.0 mL/min |
110.0 mL/min |
-- |
柱流速 |
ECD |
--
|
--
|
30.0 - 60.0 mL/min |
氮气(N2)
|