GC-2014 Detector

Gas Flow Rate


These are the standard values for the detector's gas flow rate.
Especially FPD and FTD, since the detector sensitivity is highly affected by the gas flow rate, set the appropriate flow rate by reffering the pressure-flow curve on the standard accessory card.

 
(1) For packed column

 

 

Hydrogen

Air

Makeup gas

FID

40 mL/min
(55 kPa)

400 mL/min
(40 kPa)

--

FTD

3 - 4 mL/min
(50 - 65 kPa)

80 - 160 mL/min
(30 - 60 kPa)

--

FPD
(P-mode)

130 mL/min
(135 kPa)

160 mL/min
(65 kPa)

--

FPD
(S-mode)

90 mL/min
(105 kPa)

80 mL/min
(35 kPa)

--

FPD
(Sn-mode)

90 mL/min
(105 kPa)

80 mL/min
(35 kPa)

--

ECD

--

--

2 mL/min
(10 kPa)

 

(2) For capillary column

 

 

Hydrogen

Air

Makeup gas

FID

40 mL/min
(55 kPa)

400 mL/min
(40 kPa)

30 mL/min
(75 kPa(N2),80 kPa (He))

FTD

3 - 4 mL/min
(50 - 65 kPa)

80 - 160 mL/min
(30 - 60 kPa)

30 mL/min
(80 kPa (He))

FTD2014c
(N-mode)

1 - 2 mL/min
(20 - 35 kPa)

140 - 150 mL/min
(50 - 55 kPa)

10 - 30 mL/min
(35 - 80 kPa (He))

FTD2014c
(N-mode)

3 - 6 mL/min
(50 - 95 kPa)

140 - 150 mL/min
(50 - 55 kPa)

10 - 30 mL/min
(35 - 80 kPa (He))

FPD
(P-mode)

80 mL/min
(95 kPa)

130 mL/min
(55 kPa)

--

FPD
(S-mode)

90 mL/min
(105 kPa)

80 mL/min
(35 kPa)

--

FPD
(Sn-mode)

100 mL/min
(110 kPa)

120 mL/min
(50 kPa)

--

ECD

--

--

60 mL/min
(130 kPa)

 