GC-2010Plus Detector

Gas Flow Rate


These are the standard values for the detector's gas flow rate.

 

 

Hydrogen

Air

Makeup gas
(BID: Discharge Gas, ECD-EX:ECD Gas)

 

 

FID

40.0 mL/min

400.0mL/min

30.0mL/min

 

FTD
(N-mode)

1.0 - 2.0 mL/min

140.0 - 150.0 mL/min

10.0 - 30.0 mL/min

 

FTD
(P-mode)

3.0 - 6.0 mL/min

140.0 - 150.0 mL/min

10.0 - 30.0 mL/min

 

FPD Plus
(P-mode)

62.5 mL/min

90.0 mL/min

--

 

FPD Plus
(S-mode)

40.0 mL/min

60.0 mL/min

--

 

FPD Plus
(Sn-mode)

75.0 mL/min

65.0 mL/min

--

 

ECD

--

--

30.0 - 60.0 mL/min

Nitrogen gas (N2)

ECD-EX

--

--

60.0 mL/min

Nitrogen gas (N2)

BID

 

 

Column Flow less than 10mL/min

:50mL/min

 

Column Flow 10-15mL/min

:70mL/min

 

Column Flow more than 15mL/min

:100mL/min

Helium gas (He)

 