Maintenance works must only be undertaken after the temperature of each part has been lowered to less than 50 deg.C.
Place any detached items on a clean sheet or tray. Do not lose or contaminate these items.
Undertake the works using the appropriate tools. Any dirt on the tools being used must be wiped off in advance using gauze etc. dipped in acetone.
Phillips screwdriver, tweezers, hexnut screwdriver (option, Part No. 670-18800)
1 |
Shutdown the system. |
|
2 |
|
Remove the screws of the collector holding bracket. |
3 |
|
Remove the collector unit together with the collector holding bracket.
|
4 |
|
Check for dirt inside the collector. If dirty, wipe the dirt away using cotton buds soaked in solvent. If using an item like a brush, use a soft one. The damage caused by the use of a wire brush etc. can result in noise. Replace the collector if it is so dirty that it cannot be washed, or if it rusts. Collector ASSY, FID: Part No.: 221-72322-91 |
5 |
|
Disconnect the signal cable. |
6 |
|
Pull out the high voltage plate upwards.
|
7 |
|
Pull the high voltage brush out together with the spring. |
8 |
|
Disconnect the high voltage cable. |
9 |
|
Loosen the FID jet with a hexnut screwdriver. |
10 |
|
Detach the FID jet with tweezers.
|
11 |
FID jet (0.3 dia. standard) 221-48258-91 |
Check for any dirt or blockages in the jet. If dirty, wipe the dirt away using a cotton bud soaked in solvent. Replace it with a new item if it has blockages. Note: |
12 |
|
Set the FID jet and use the nut driver to turn it until it comes to a firm stop. |
13 |
Put back the high voltage unit and collector unit by following the removal procedure in reverse. |
|
14 |
Start up the system. |
Maintenance Top ; GC-2010 Main ; GC-2010 Detector ; GC-2010 Sample Injection Port ; GC-2010 Column Oven